Specifications
SKU: FLUKE-5560A
Fluke 5560A High-Performance Multi-Product Calibrator
The Fluke 5560A Multi-Product Calibrator provides a versatile source for calibrating electrical test instruments in calibration laboratories and manufacturing environments.
Workload Coverage And Calibration Functions
The 5560A is designed for demanding electrical calibration workloads, including popular 6.5 digit bench multimeters at a 4:1 test uncertainty ratio for the cited workloads. Check the complete uncertainty budget for each calibration point. Its synthesised inductance and continuous high-current output extend coverage beyond basic voltage and resistance checks.
Operation And Automated Calibration
A colour touchscreen and illuminated Visual Connection Management terminals guide setup. MET/CAL compatibility supports automated procedures. Phase-adjustable AC power, power factor control, dual voltage outputs and simple harmonics support additional electrical workloads.
Oscilloscope And High-Current Options
Scope options provide amplitude, fast-edge, bandwidth, time-marker, impedance and overload checks. External 52120A amplifier configurations can provide higher current; appropriate current coils extend effective current for clamp meters and Rogowski coils. These accessories are separate configuration items.
Key Features
- AC and DC voltage and current calibration
- Resistance, capacitance and inductance functions
- Temperature signal simulation for compatible instruments
- Optional oscilloscope calibration packages
- 30 A continuous output without a duty-cycle restriction
- Support for 17 thermocouple types and 10 RTD types
- Multi-unit synchronisation and isolated connection terminals
- Effective current display for compatible Fluke coils
Technical Specifications
| Specification | Details |
|---|---|
| Model | 5560A |
| Current Output | Up to 30.2 A |
| Optional Oscilloscope Bandwidth | 600 MHz, 1 GHz or 2 GHz, depending on option |
| DC Voltage | 0 to ±1,020 V |
| DC Current | 0 to ±30.2 A continuous |
| AC Voltage | 1 mV to 1,020 V; 3 Hz to 500 kHz, subject to range limits |
| AC Current | 10 µA to 30.2 A; 3 Hz to 30 kHz, subject to range limits |
| Variable Resistance | 0 Ω to 1,200 MΩ |
| Variable Capacitance | 0 pF to 120 mF |
| Variable Inductance | 13 µH to 120 H |
| Temperature Simulation | 17 thermocouple types; 10 RTD types; ITS-90 or IPTS-68 |
| Scope Option Fast Edge | Less than 175 ps |
| Scope Time-Marker Accuracy | 2.5 ppm |
| External Current Expansion | Up to 360 A with amplifier configuration; up to 6,000 A effective current with suitable amplifier/coil configuration |
Applications
Electrical calibration laboratories, calibration of bench and handheld multimeters, clamp-meter verification, temperature signal simulation and optional oscilloscope calibration.
Configuration And Selection Notes
Ranges are summary limits, not simultaneous output guarantees. Accuracy varies with function, range, frequency, temperature and calibration interval. Scope packages, current amplifiers, coils and software must be specified separately.
Contact us to enquire about the Fluke 5560A Multi-Product Calibrator and the configuration or accessories suited to your application.
Inquiry - Fluke 5560A High-Performance Multi-Product Calibrator