Specifications
SKU: 6270A
Fluke 6270A Pressure Controller And Calibrator
The Fluke 6270A Pressure Controller And Calibrator supports automated pneumatic pressure calibration for pressure instruments and sensors.
Configure Ranges Around Your Workload
The 6270A can mix different module ranges and accuracy classes in one chassis. This lets a laboratory allocate higher performance to demanding ranges while using other modules for dial gauges, switches or lower-accuracy transmitters. Modules can be added as the workload changes.
Three Pressure Measurement Module Families
PM200 modules use silicon sensors with a 0.02% full-scale specification. PM500 modules offer 0.01% of reading from 50% to 100% of span for most ranges. PM600 modules use quartz reference pressure transducer technology and offer 0.01% of reading from 30% to 100% of span. Consult the module specification for the selected range and mode.
Maintenance, Protection And Automation
Measurement and control modules are accessible from the front, including in rack installations. Separate modules simplify servicing and range changes. Pressure relief valves, configurable pressure limits and an emergency abort control provide protection. Optional contamination prevention helps protect the system from residual liquids and particles in devices under test. COMPASS for Pressure software can automate calibration sequences on one or multiple devices.
Key Features
- Automated pneumatic pressure control
- Accommodates up to five pressure measurement modules
- Supports PM200, PM500 and PM600 module families
- Configurable for different pressure calibration workloads
- Front-access module replacement without specialised tools
- Modules can be calibrated in the chassis or with an optional external calibration kit
- Remote interface for custom calibration software
- Optional integration with PG7000-series piston gauge systems
Technical Specifications
| Specification | Details |
|---|---|
| Model | 6270A |
| Maximum Pressure | Up to 20 MPa (3,000 psi), configuration dependent |
| Measurement Modules | Up to 5 |
| Pressure Coverage | Low differential pressure to 20 MPa; vacuum capability depends on configuration |
| PM200 Performance | 0.02% of full scale |
| PM500 Performance | 0.01% of reading over 50–100% of span for most ranges |
| PM600 Performance | 0.01% of reading over 30–100% of span |
| Rear Connection Options | NPT, BSP or 7/16-20 manifold configurations |
Applications
Calibration laboratories, pressure sensor manufacturing, gauge and transmitter calibration, pressure switch testing and automated pressure test systems.
Configuration And Selection Notes
The chassis does not establish a universal accuracy or pressure range. Select measurement modules, the control module, pressure supply, connection manifold and protection accessories for the intended devices.
Contact us to enquire about the Fluke 6270A Pressure Controller And Calibrator and the configuration or accessories suited to your application.
Inquiry - Fluke 6270A Pressure Controller / Calibrator